Somru BioScience

Ranibizumab(Lucentis®) PK

$0.00

Catalog #: SB-07-050

SKU: SB-07-050 Category: Tag:
Category

Pharmacokinetic-(PK)

Species

Size

1 pack ( 2 x 96 wells)

Data Sheet

Ranibizumab (Eylea®) – PK ELISA

Catalogue Number: SB-07-050

For the quantitative determination of Ranibizumab in human serum and plasma

Notice: This product specification data sheet is for information purposes only. Please contact tec@somrubioscience.com for up to date information.

Product Specifications Data Sheet

About The Somru® Ranibizumab ELISA kit is designed to measure ranibizumab with high specificity and enhanced sensitivity in serum or plasma. This user-friendly assay design utilizes biotinylated target protein (VEGF) to capture ranibizumab and anti-ranibizumab polyclonal antibody for detection.
Detection Method Colorimetric; absorbance at 450 nm with reference absorbance at 650 nm
Storage Kit ships at 2 to 8°C. Multiple storage temperatures upon arrival.
Calibrator Ranibizumab
Calibration Curve 250 to 10,000 pg/mL; 4 or 5 -parameter curve fit
Assay Range 250 to 5,000 pg/mL
Sensitivity 250 pg/mL (Please contact tec@somrubioscience.com for high sensitivity method)
Precision Intra-assay precision ≤ 10%. Inter-assay mean precision ≤ 15%. Determined by analyzing three levels of quality control samples in 6 replicates on 6 different occasions.
Accuracy The Mean %bias is ≤ 15%. Determined by analyzing three levels of quality control samples in 6 replicates on 6 different occasions.
Total Error The Total Error is ≤ 20%. Calculated by adding the mean %bias with inter-assay variability (%CV).
Safety Warnings and Precautions All reagents and samples should be handled as if potentially hazardous and infectious. It is recommended that all kit reagents be handled by trained laboratory personnel and are used in accordance with the principles of Good Laboratory Practice.

 

Notice: Somru® ELISA kits are custom-made to order. Each kit lot is tested to ensure that it meets specifications before shipment. Minor adjustments to the preparation of reagents may occur due to lot-to-lot variation. Please always refer to the product sheet enclosed with each kit before performing each experiment.

                FOR RESEARCH USE ONLY

Please Login To Download Attachment