Data Sheet
Aflibercept (Eylea®) – PK ELISA
Catalogue Number: SB-07-089
For the quantitative determination of Aflibercept in human serum and plasma
Notice: This product specification data sheet is for information purposes only. Please contact tec@somrubioscience.com for up to date information.
Product Specifications Data Sheet
About | The Somru® Aflibercept ELISA kit is designed to measure aflibercept with high specificity and enhanced sensitivity in serum or plasma. This user-friendly assay design utilizes target (VEGF) to capture aflibercept and anti-aflibercpt antibody for detection. |
Detection Method | Colorimetric or Luminescence |
Storage | Kit ships at 2 to 8°C. Multiple storage temperatures upon arrival. |
Calibrator | Eylea |
Calibration Curve | 25 to 0.78 ng/mL; 4 or 5 parameter curve fit with appropriate weighting |
Sensitivity | 0.78 ng/mL |
Precision | Intra-assay precision ≤ 10%. Inter-assay mean precision ≤ 15%. Determined by analyzing three levels of quality control samples in 6 replicates on 6 different occasions. |
Accuracy | The Mean %bias is ≤ 15%. Determined by analyzing three levels of quality control samples in 6 replicates on 6 different occasions. |
Total Error | The Total Error is ≤ 20%. Calculated by adding the mean %bias with inter-assay variability (%CV). |
Safety Warnings and Precautions | All reagents and samples should be handled as if potentially hazardous and infectious. It is recommended that all kit reagents be handled by trained laboratory personnel and are used in accordance with the principles of Good Laboratory Practice. |
Notice: Somru® ELISA kits are custom-made to order. Each kit lot is tested to ensure that it meets specifications before shipment. Minor adjustments to the preparation of reagents may occur due to lot-to-lot variation. Please always refer to the product sheet enclosed with each kit before performing each experiment.
FOR RESEARCH USE ONLY